Detection CMOS Cameras

The detection system consists of camera, lens and scintillator screens. Below we list of the camera in use at 32-ID:

Camera

Web Site

pixels (HxV)

pixels size (μm)

bit

fps

Manual

Part number

Purchase order

Server

Blackfly S GigE

flir_web_site

5230 x 3032

2.74

8-10-12

7

camera_specs

BFS-PGE-161S7M-C

camera_order_00001

gauss

Teledyne Oryx

flir_web_site

6464 x 4852

3.45

8-10-12

26

camera_specs2

Oryx 10GigE: 31 MP

camera_order_00002

gauss

Teledyne Kinetix

flir_web_site

3200 x 3200

6.5

8-10-12

500

camera_specs3

Kinetix

camera_order_00003

maxwell

To start/stop/check the detector (ORYX) epics IOC - mind to use the correct server:

[usertxm@gauss]$ 32idbSP1 start
[usertxm@gauss]$ 32idbSP1 stop
[usertxm@gauss]$ 32idbSP1 status

the detector user interface is accessible with:

[usertxm@gauss]$ 32idbSP1 medm

To start/stop/check the detector (KINETIX) epics IOC:

[usertxm@maxwell]$ 32kinetix start
[usertxm@maxwell]$ 32kinetix stop
[usertxm@maxwell]$ 32kinetix status

the detector user interface is accessible with:

[usertxm@maxwell]$ 32kinetix medm

or by selecting detector from the main TXM medm screen.

Note: old version of AreaDetector for the TXM camera is accessible using 32idARV1 alias.

Power cycle

To power cycle the Blackfly you can use the power management unit (PDI) s32pdu1 outlet #1